6243/44 DC Voltage Current Source/Monitor Operation Manual
8.2.6 Initializing Calibration Data
8-17
(4) Exiting calibration
NOTE: Upon completion of calibration, be sure to set Cal Switch Set to OFF.
1. Press the button on the calibration item selection screen.
The following screen appears asking whether or not the data is to be saved.
2. To save the data, press the 1 button. - or -
To exit without saving the data, press the 0 button.
The standby status is invoked and the parameter selection screen appears.
3. Select the Cal Switch Set parameter and set it to OFF.
4. Press the EXIT button.
The normal measurement screen returns.
This completes calibration.
8.2.6 Initializing Calibration Data
Normally, the calibration data is not initialized unless the calibration values are initialized.
NOTE: When the calibration data is initialized and saved, the default calibration values are saved and the
previous calibration data are lost.
Be sure to recalibrate all the ranges in this case.
1. Select Init to Default on the Calibration Item Selection screen in the calibration
mode and press the button.
2. Press the DIRECT button.
Upon completion of calibration data initialization, “Done” appears.
AZERO
Save?
Cancel :’0’ Save :’1’
Init to Default
Execute (Press DIRECT Key)