6243/44 DC Voltage Current Source/Monitor Operation Manual
9. SPECIFICATIONS
9-1
9. SPECIFICATIONS
This chapter describes the specifications of 6243/44.
All accuracy specifications are guaranteed for one year at a temperature of 23 5C and a relative humidity
not exceeding 85%.
9.1 6243 Specifications
9.1.1 Voltage/Current Source
Source range:
Voltage, current limiters (compliance) range:
Range Source Range Resolution
Voltage Source 320 mV
3.2 V
32 V
110 V
0 to 320.00 mV
0 to 3.2000 V
0 to 32.000 V
0 to 110.00 V
10 V
100 V
1 mV
10 mV
Current source 32 A
320 A
3.2 mA
32 mA
320 mA
2 A
0 to 32.000 A
0 to 320.00 A
0 to 3.2000 mA
0 to 32.000 mA
0 to 320.00 mA
0 to 2000.0 mA
1 nA
10 nA
100 nA
1 A
10 A
100 A
Range
Maximum
setting range
Minimum
setting range
Resolution
Voltage limiter 320 mV
3.2 V
32 V
110 V
320.00 mV
3.2000 V
32.000 V
110.00 V
3 mV
30 mV
300 mV
3 V
10 V
100 V
1 mV
10 mV
Current limiter 32 A
320 A
3.2 mA
32 mA
320 mA
2 A
32.000 A
320.00 A
3.2000 mA
32.000 mA
320.00 mA
2000.0 mA
300 nA
3 A
30 A
30
0 A
3 mA
30 mA
1 nA
10 nA
100 nA
1 A
10 A
100 A