6243/44 DC Voltage Current Source/Monitor Operation Manual
5.2.10 GPIB Code List
5-22
SOURCE
(Continued)
Source function and
source range settings
V?/I? he source function and source range query command
Responds with V or I...
Source value (pulse
value) and limiter
value
(Continued)
D Unit specification: The optimal range is set automatically.
D<data> UNIT
Units: MV, V, UA, MA, and A
The limiter value is set when a unit that is not used for the current source function is spec-
ified.
No unit specification: The current source function and source range are used.
D<data>
Default value: D+000.00E-3V, D0500.0E-3A (6243)
D+000.00E-3V, D04.000E+0A (6244)
D? The source value and limiter value query command
Responds with: D  aaaaaaE+aUNIT, DbbbbbbE+bUNIT
aaaa: Voltage or current source value
bbbb: Current or voltage limiter value (polarity is indicated by space.)
Unit: V or A
The voltage and current value responses depend on the current source function.
Pulse base value DB Unit specification: The optimal range is set automatically.
DB<data>UNIT The <data> unit is ignored.
Units: MV, V, UA, and MA
An error occurs when a unit that is not used for the current source function is specified,
Default value (6243): DB 000.00E-3V (for the voltage function)
DB 00.000E-6A (for the current function)
Default value (6244): DB 000.00E-3V (for the voltage function)
DB 000.00E-6A (for the current function)
DB? The pulse base value query command.
Responds with: DB aaaaaaE+aUNIT
aaaa: Voltage or current source value (polarity is indicated by space.)
Unit: V or A
The voltage and current base value responses depend on the current source function.
Limiter polarity
mode
PL0 Auto: Enters the same polarity as the source value.
PL1 Plus: Enters the correct positive value.
PL2 Minus: Enters the correct negative value.
PL? The limiter polarity mode query command.
Responds with PL...
Remote sensing RS0 2 Wire
RS1 4 Wire
RS? The setting remote sensing query command.
Responds with RS...
The oscillation,
reverse polarity
source, and limiter
detection buzzer
UZ0 OFF
UZ1 The buzzer sounds when the oscillation, reverse polarity source, or limiter is detected.
UZ? The oscillation, reverse polarity source, and limiter detection buzzer query command.
Responds with UZ...
Buffer control B The source value can be buffer controlled and output using *TRG.
The H command sets the buffer control to OFF. (An error occurs when the DC mode is not
used.)
OFF *1
B? Enquires about the buffer controlled source value.
The same response as the source value query response is used. (Without the limiter value.)
When the buffer is set to OFF, D+999.99E+9V or D+999.99E+9A.
OPERATE Operating or standby E Output is set to ON. (for the operating status)
H Output is set to OFF. (for the standby status) *1
E? The output ON or OFF query command.
H? Responds with "E" and "H."
MEASURE Measurement
function
F0 Not measured.
F1 Voltage measurement (VM)
F2 Current measurement (IM)
F? The measurement function query command.
Responds with F...
Auto range R0 Auto range
R1 The limiter value range is used.
R? The auto range query command
Responds with R...
Table 5-5 GPIB Command List (2 of 9)
Item Command Description Default
Operation
During DC/
PLS operation
During sweep
operation