6243/44 DC Voltage Current Source/Monitor Operation Manual
5.2.11 Program Examples
5-40
5.2.11.5 Program Example 5: Using Measurement Buffer Memory
(100 measurement data is recalled in the shortest time.)
Option Explicit Explicit declaration for all variables
Private Sub Start_Click() Event procedure for the command button (Start)
Dim board As Integer GPIB board address
Dim pad As Integer 6243/44 address
Dim vig As Integer 6243/44 device descriptor
Dim dt As String * 20 Data reception buffer
Dim dt_sz As Integer Number of measurement buffer memory data
Dim dt_rn(100) As String * 15 Measurement buffer memory data storage string variable
Dim i As Integer, s As Integer i: Loop variable, s: Serial poll result storage variable
board = 0 GPIB board address 0
pad = 1 6243/44 address 1
Call ibdev(board, pad, 0, T30s, 1, 1, vig) Opening and initializing device (6243/44) (timeout 30 s)
Call ibconfig(vig, IbcUnAddr, 1) Address setting performed for each transmission or reception
Call ibeos(vig, &H40A) Setting terminator to LF for sending command
Executing sweep measurement
Call SUBsend(vig, “C,*RST”) DCL and parameter initialization
Call SUBsend(vig, “*CLS”) Status byte initialization
Call SUBsend(vig, “*SRE8”) Setting bit 3 for the Service Request Enable Register to 1
Call SUBsend(vig, “DSE8192”) Setting bit 13 for the Device Event Enable Register to 1
Call SUBsend(vig, “S0”) SRQ transmission mode
Registering setting for transmitting SRQ following completion of sweep
Call SUBsend(vig, “MD2”) Sweep generation mode
Call SUBsend(vig, “SN0.1V,10V,0.1V”) Linear sweep: Start 0.1 V, stop 10 V, step 0.1 V
Call SUBsend(vig, “SB0V”) Sweep bias value 0 V
Call SUBsend(vig, “SP3,4,100”) Hold time 3 ms, measure delay time 4 ms
Period 100 ms
Call SUBsend(vig, “D30MA”) Limiter value 30 mA
Call SUBsend(vig, “SM1”) Memory Store ON
Call SUBsend(vig, “E”) Output ON
Call SUBsend(vig, “*TRG”) Starting sweep
Waiting for sweep measurement completion
Call ibwait(vig, RQS Or TIMO) Waiting for SQL transmission
If (ibsta And TIMO) Then In case of timeout
Call MsgBox(“SRQ Time Out”, vbOKOnly, “Error”)
indicating error