6243/44 DC Voltage Current Source/Monitor Operation Manual
9.2.1 Voltage/Current Source
9-16
Settling time; The time to reach, from the start time, 0.03% of the final value, by changing the out-
put from zero to full scale. However, in the case of a purity load resistor, load capac-
ity will be below 2.5 pF, source value, and limiter setting will be at full scale.
Line regulation: 0.003% of range
Load regulation: Voltage source: Within 0.003% of range at the maximum load and with 4 Wire
connection
Current source: Depends on CMV section (A Vo/1 V) of over all accuracy.
Output resistance: 2 Wire connection excluding output cable
Maximum load capacity:
Maximum load capacity at which oscillation does not occur during voltage source or
when the voltage limiter is operating.
Maximum inductance: Maximum inductance at which oscillation does not occur during current
source or when the current limiter is operating.
Range Settling time
Voltage source 320 mV
300 s or less
3.2 V
20 V 700 s or less
Current source 320 A
3 ms or less
3.2 mA
32 mA
320 mA
3.2 A
10 A
Range
Output resistance
Maximum load capacity
Voltage source Current source
320 A
3.2 mA
32 mA
320 mA
3.2 A
10 A
100 m or lower
10 m or lower
10 m or lower
10 m or lower
10 m or lower
10 m or lower
1 10
9
or higher
1 10
8
or higher
1 10
7
or higher
1 10
6
or higher
1 10
5
or higher
1 10
4
or higher
1 F
100 F
100 F
2000 F
2000 F
2000 F
Resistance of standard accessory cable: 100 m or lower
Current source range
Current limiter range
Maximum inductance
320 A 500 H
3.2 mA to 10 A 1 mH

Table of Contents