6243/44 DC Voltage Current Source/Monitor Operation Manual
9.2.2 Voltage/Current Measurement
9-19
Effective CMRR: Unbalanced impedance 1 k
DC and AC 50/60 Hz 0.08%
NMRR: AC 50/60 Hz 0.08%
Integration time
500 s to 10 ms 1 PLC to 100 PLC
Voltage measurement/
Current measurement
60 dB 120 dB
Integration time
500 s to 10 ms 1 PLC to 100 PLC
Voltage measurement/
Current measurement
0 dB 60 dB