6243/44 DC Voltage Current Source/Monitor Operation Manual
2.3.1 Measuring a Diode
2-33
Acquiring NULL data (canceling errors caused by cable resistance)
17. Short-curcuit the HI OUTPUT cable (red) with the LO OUTPUT lead (black).
18. Press the OPERATE key.
19. Press the NULL key.
20. Press the OPERATE key.
The status is change to Standby.
21. Connect the HI OUTPUT (red) lead to the diode’s anode.
22. Press the OPERATE key.
The measured result of the diode’s forward voltage appears.
23. Press the NULL and OPERATE keys.
The status is change to Standby.
Measuring the diode reverse leak current (IR)
24. Press the MODE key 3 times.
The mode becomes the DC source.
25. Press the VS/IS key.
The function becomes the Voltage source.
26. Press the VM/IM key.
The function becomes the Current measurement.
AUTO AZERO
VM+007.654mV
IS:+100.00mA L: 1.5000V *PLS
AUTO NULL AZERO
VM+000.006mV
IS:+100.00mA L: 1.5000V *PLS
AUTO NULL AZERO
VM+0.77687V
IS:+100.00mA L: 1.5000V *PLS