7352 Series Digital Multimeter Operation Manual
6.6.3 ADC Command Reference
6-26
Calculation Calculation
between 2
measurements
MCLn Calculation between 2 measurements
Performs four arithmetic operations between measurement
values for the right and left side displays.
n: 0: OFF
1: Measurement for the left side display + Measure-
ment for the right side display
2: Measurement for the left side display - Measure-
ment for the right side display
3: Measurement for the left side display * Measure-
ment for the right side display
4: Measurement for the left side display / Measure-
ment for the right side display
NOTE:
Valid only when the 2nd display is ON.
0–
MCL? Reply: MCLd –
System 2nd display DE0 2nd display OFF –
DE1 2nd display ON
DE? Reply: DE0 or DE1 –
1st/2nd
display
selection *5
DSP1 Selects the 1st display. –
DSP2 Selects the 2nd display.
NOTE:
Valid in a line where this command is set.
The 1st display is selected in a line where this
command is omitted.
–
Output data SD0 Remotely output the 1st and 2nd displays.
SD1 Remotely output the 1st display only.
SD2 Remotely output the 2nd display only.
NOTE:
When the 2nd display is OFF, this command is
changed to SD0.
SD? Reply: One of SD0 to SD2 –
Buzzer BZ0 OFF BZ0
BZ1
BZ2
BZ3
BZ4
BZ1 ON
BZ? Reply: BZ0 or BZ1 –
*5: This command cannot be used when the calibration mode is set to ON.
Item
Command
Description
Va l i d
command
for the 2nd
display
setting
Initial
value
R6452