7352 Series Digital Multimeter Operation Manual
6.6.4 SCPI Command Reference
6-34
Trigger
system
Start :INITiate Exits the IDLE state and moves to the trigger waiting state.
Continue :INITiate:CONTinuous
{ON,1|OFF,0}
ON, 1: CONTINUOUS ON
OFF, 0: CONTINUOUS OFF
:INITiate:CONTinuous? Reply: 1 or 0
Trigger abort :ABORt Forcibly moves to the IDLE state.
Trigger source
selection
:TRIGger:SOURce
{IMMediate|MANual|
EXTernal|BUS}
IMMediate
MANual
EXTernal (7352A)
BUS
:TRIGger:SOURce? Reply: IMM, MAN, EXT, or BUS
Trigger delay :TRIGger:DELay
{<real>|MAXimum|MINimum}
[,(@<ch>)] *2
<real>: 0 to 3600 (seconds) (0)
:TRIGger:DELay?
[{MAXimum|MINimum}]
[,(@<ch>)] *2
Reply: +d.ddddEdd
Each trigger
delay for the
right and left
displays
:TRIGger:DELay:EACH
{ON,1|OFF,0}
ON, 1: Each trigger delay ON
OFF, 0: Each trigger delay OFF
:TRIGger:DELay:EACH? Reply: 1 or 0
Trigger count :TRIGger:COUNt
{<real>|MAXimum|MINimum}
<real>: 1 to 50000
:TRIGger:COUNt?
[{MAXimum|MINimum}]
Reply: ddddd
Sampling count
(1 trigger)
:SAMPle:COUNt
{<int>|MAXimum|MINimum}
<int>: 1 to 16000
:SAMPle:COUNt?
[{MAXimum|MINimum}]
Reply: ddddd
*2: [,(@<ch>)] in the trigger delay can only be set when each trigger delay for the right and left displays is set to ON.
Item SCPI command Description