EasyManuals Logo

ADCMT 7352A User Manual

ADCMT 7352A
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #138 background imageLoading...
Page #138 background image
7352 Series Digital Multimeter Operation Manual
6.6.4 SCPI Command Reference
6-34
Trigger
system
Start :INITiate Exits the IDLE state and moves to the trigger waiting state.
Continue :INITiate:CONTinuous
{ON,1|OFF,0}
ON, 1: CONTINUOUS ON
OFF, 0: CONTINUOUS OFF
:INITiate:CONTinuous? Reply: 1 or 0
Trigger abort :ABORt Forcibly moves to the IDLE state.
Trigger source
selection
:TRIGger:SOURce
{IMMediate|MANual|
EXTernal|BUS}
IMMediate
MANual
EXTernal (7352A)
BUS
:TRIGger:SOURce? Reply: IMM, MAN, EXT, or BUS
Trigger delay :TRIGger:DELay
{<real>|MAXimum|MINimum}
[,(@<ch>)] *2
<real>: 0 to 3600 (seconds) (0)
:TRIGger:DELay?
[{MAXimum|MINimum}]
[,(@<ch>)] *2
Reply: +d.ddddEdd
Each trigger
delay for the
right and left
displays
:TRIGger:DELay:EACH
{ON,1|OFF,0}
ON, 1: Each trigger delay ON
OFF, 0: Each trigger delay OFF
:TRIGger:DELay:EACH? Reply: 1 or 0
Trigger count :TRIGger:COUNt
{<real>|MAXimum|MINimum}
<real>: 1 to 50000
:TRIGger:COUNt?
[{MAXimum|MINimum}]
Reply: ddddd
Sampling count
(1 trigger)
:SAMPle:COUNt
{<int>|MAXimum|MINimum}
<int>: 1 to 16000
:SAMPle:COUNt?
[{MAXimum|MINimum}]
Reply: ddddd
*2: [,(@<ch>)] in the trigger delay can only be set when each trigger delay for the right and left displays is set to ON.
Item SCPI command Description
Initial
value

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADCMT 7352A and is the answer not in the manual?

ADCMT 7352A Specifications

General IconGeneral
BrandADCMT
Model7352A
CategoryMultimeter
LanguageEnglish

Related product manuals