10 Agilent 1200 Infinity Series DAD User Manual
1 Introduction
Overview of the Module
Overview of the Module
The detector is designed for highest optical performance, GLP compliance and
easy maintenance. It includes the following features:
• Maximum of 160 Hz (G4212A) or 80 Hz (G4212B) data acquisition rate.
• Higher sensitivity for conventional LC as well as ultra fast applications by
using next generation optical design.
• Increased sensitivity with 60 mm Max-Light cartridge flow cell.
• Optimized cell geometry for less peak dispersion for narrow bore
applications.
• Max-Light cartridge flow cells for standard and bio-inert applications are
available.
• More reliable and robust peak integration process (automated) due to less
baseline noise/drift/refractive index and thermal effects especially under
ultra fast gradient conditions.
• RFID tracking technology is used for the UV-lamp and the Max-Light
cartridge flow cells.
• Multiple wavelength and full spectral detection at 160 Hz (G4212A)/80 Hz
(G4212B) sampling rate, keeping up with the analysis speed of ultra-fast
LC.
• Programmable 1 – 8 nm slit (G4212A) or fixed 4 nm slit (G4212B) for rapid
optimization of sensitivity, linearity and spectral resolution provides
optimum incident light conditions.
• Improved Electronic temperature control (ETC) provides maximum
baseline stability and practical sensitivity under fluctuating ambient
temperature and humidity conditions.
• Additional diagnostic signals for temperature and lamp voltage monitoring.
• Easy exchange of flow cell by cartridge design.
For specifications, see “Performance Specifications G4212A” on page 24 or
“Performance Specifications G4212B” on page 26.