56 Agilent 1200 Infinity Series DAD User Manual
4 Using the Module
Setting up the Detector with Agilent ChemStation
Control Settings
Lamps: can be turned ON/OFF.
At Power On: automatic lamp-on at power on.
Analog Output Range: can be set to either 100 mV or 1 V
full scale (1 V = default).
UV lamp Tag: Automatic detects a lamp with RFID tag. If
no RFID tag lamp is used, “UV lamp not ready” is
displayed and it cannot be ignited. A compatible mode has
to be selected based on the used lamp; see Non-RFID-tag
lamp information below.
Automatic Turn On: The module can be turned on at a
specified date/time. If “Turn UV lamp on” at power on is
set, the lamp is turned on as well.
Non-RFID-tag lamp
In case a non-RFID-tag lamp is used, the user interface
will show this when selecting a compatible mode.
You may operate the detector outside of the guaranteed
specification.