Agilent 1200 Infinity Series DAD User Manual 179
Maintenance
9
Storage of Max-Light Cartridge Cell
Storage of Max-Light Cartridge Cell
1 Flush the Max-Light Cartridge Flow Cell with iso-propanol or methanol and
insert the plugs into the cell inlet and outlet (see “Replacing the Max-Light
Cartridge Cell” on page 173).
2 Remove the Max-Light Cartridge Cell from the cartridge holder of the
detector.
3 Replace the black hoods, that secure the cell light inlet and outlet.
4 Store the Max-Light Cartridge Cell in plastic case provided with the
Max-Light Cartridge Flow Cell.