126 Agilent 1200 Infinity Series DAD User Manual
7 Error Information
Detector Error Messages
Detector Error Messages
These errors are detector specific.
Diode Current Leakage
Error ID: 1041
When the detector is switched on, the processor checks the leakage current of
each of the optical diodes. If the leakage current exceeds the upper limit, the
error message is generated.
UV Lamp Current
Error ID: 7450
The UV lamp current is missing.
The processor continually monitors the anode current drawn by the lamp
during operation. If the anode current falls below the lower current limit, the
error message is generated.
Probable cause Suggested actions
1 Defective PDA/optical unit.
Please contact your Agilent service
representative.
2 Defective connector or cable.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Lamp disconnected.
Ensure the UV lamp connector is seated firmly.
2 Defective UV lamp or non-Agilent lamp.
Exchange the UV lamp.
3 Defective detector main board.
Please contact your Agilent service representative.
4 Defective power supply.
Please contact your Agilent service representative.

Table of Contents

Other manuals for Agilent Technologies 1200 Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 Infinity Series and is the answer not in the manual?

Agilent Technologies 1200 Infinity Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 Infinity Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals