134 Agilent 1200 Infinity Series DAD User Manual
8 Test Functions and Calibration
Introduction
Introduction
All tests are described based on the Agilent Lab Advisor Software B.01.03.
Other user interfaces may not provide any test or just a few.
For details on the use of the interface refer to the interface documentation.
Tabl e 1 0 Interfaces and available test functions
Interface Comment Available Function
Agilent Instrument Utilities Maintenance tests are
available
• Intensity
• Cell
• WL Calibration
Agilent Lab Advisor All tests are available • Self-Test
• Intensity
• Quick Noise
• ASTM Drift and Noise
• Cell
• Dark Current
• D/A Converter
• Slit (G4212A only)
• WL Verification
• WL Calibration
• Test Chromatogram (Tools)
• Spectra Scan (Tools)
• Module Infos (Tools)
• Diagnostic (Tools)
Agilent ChemStation No tests available
Adding of temperature/lamp
signals to chromatographic
signals possible
• Temperature main board
• Temperature optical unit
• Lamp anode voltage
Agilent Instant Pilot Some tests are available • Intensity
• WL Calibration
• Cell