EasyManua.ls Logo

Agilent Technologies 1200 Infinity Series - Heater Power at Limit

Agilent Technologies 1200 Infinity Series
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1200 Infinity Series DAD User Manual 131
Error Information
7
Detector Error Messages
Heater Power At Limit
Error ID: 1074
The available power of the heater reached either the upper or lower limit. This
event is sent only once per run. The parameter determines which limit has
been hit:
0 means upper power limit hit (excessive ambient temperature drop).
1 means lower power limit hit (excessive ambient temperature increase).
Probable cause Suggested actions
1 Excessive ambient temperature change.
Wait until temperature control equilibrates.