Agilent 1200 Infinity Series DAD User Manual 129
Error Information
7
Detector Error Messages
Wavelength Recalibration Lost
Error ID: 1037
The calibration information needed for your detector to operate correctly has
been lost.
During calibration of the detector the calibration values are stored in ROM. If
no data is available in the spectrometer ROM, the error message is generated.
Illegal Temperature Value from Sensor on Main Board
Error ID: 1071
This temperature sensor (located on the detector main board) delivered a
value outside the allowed range. The parameter of this event equals the
measured temperature in 1/100 centigrade. As a result the temperature
control is switched off.
Probable cause Suggested actions
1 The detector is new.
Recalibrate the detector.
2 The detector has been repaired.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Defective sensor or main board.
Please contact your Agilent service
representative.
2 Detector is exposed to illegal ambient
conditions.
Verify that the ambient conditions are within
the allowed range.