130 Agilent 1200 Infinity Series DAD User Manual
7 Error Information
Detector Error Messages
Illegal Temperature Value from Sensor at Fan Assembly
Error ID: 1072
This temperature sensor (located close to the fan) delivered a value outside
the allowed range. The parameter of this event equals the measured
temperature in 1/100 centigrade. As a result the temperature control is
switched off.
Heater at fan assembly failed
Error ID: 1073
Every time the deuterium lamp or the tungsten lamp (DAD only) is switched
on or off a heater self-test is performed. If the test fails an error event is
created. As a result the temperature control is switched off.
Probable cause Suggested actions
1 The temperature sensor is defect.
Please contact your Agilent service
representative.
2 Defective main board.
Please contact your Agilent service
representative.
3 Detector is exposed to illegal ambient
conditions.
Verify that the ambient conditions are within
the allowed range.
Probable cause Suggested actions
1 Defective connector or cable.
Please contact your Agilent service
representative.
2 Defective heater.
Please contact your Agilent service
representative.