232 Agilent 1200 Infinity Series DAD User Manual
13 LAN Configuration
Dynamic Host Configuration Protocol (DHCP)
Setup (DHCP)
1 Note the MAC address of the LAN interface (provided with G1369C LAN
Interface Card or Main Board). This MAC address is on a label on the card
or at the rear of the main board, e.g. 0030d3177321.
On the Instant Pilot the MAC address can be found under Details in the LAN
section.
Figure 63 LAN Setting on Instant Pilot
2 Set the Configuration Switch to DHCP either on the G1369C LAN Interface
Card or the main board of above mentioned modules.
Software required The modules in the stack must have at least firmware from set A.06.34 and the above mentioned
modules B.06.40 or above (must from the same firmware set).
Tabl e 2 4 G1369C LAN Interface Card (configuration switch on the card)
SW 4 SW 5 SW 6 SW 7 SW 8 Initialization Mode
ON OFF OFF OFF OFF DHCP

Table of Contents

Other manuals for Agilent Technologies 1200 Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 Infinity Series and is the answer not in the manual?

Agilent Technologies 1200 Infinity Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 Infinity Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals