180 Agilent 1200 Infinity Series DAD User Manual
9 Maintenance
Drying the Leak Sensor
Drying the Leak Sensor
When If leak has occurred.
Tools required Description
Tissue
Preparations Turn the pump off.
1 Remove the front cover. 2 Locate the leak sensor area.
3 Dry the leak sensor and the area around. Check loose
fittings at the flow cell. Note that the leak sensor does
not touch the panel (clearance of about 1 mm).
4 Close the front cover.