Replacing the Max-Light Cartridge Cell
Replacing the Max-Light Cartridge Cell
For bio-inert modules use bio-inert parts only!
When If leaking or if intensity drops due to contaminated flow cell.
Tools required Description
Wrench. hexagonal
Parts required # p/n Description
1 G4212-60008
Max-Light Cartridge Cell (10 mm, V(
σ) 1.0 µL)
1 G4212-60007
Max-Light Cartridge Cell (60 mm, V(
σ) 4.0 µL)
1 G4212-60011 Max-Light Cartridge Test Cell
1 G5615-60018
Max-Light Cartridge Cell Bio-inert (10 mm, V(
σ) 1.0 µL)
includes Peek Capillary 1.5 m i.d. 0.18 mm (0890-1763) and PEEK
Fittings 10/PK (5063-6591)
1 G5615-60017
Max-Light Cartridge Cell Bio-inert (60 mm, V(
σ) 4.0 µL)
includes Peek Capillary 1.5 m i.d. 0.18 mm (0890-1763) and PEEK
Fittings 10/PK (5063-6591)
1 G4212-60032
HDR Max-Light Cartridge Cell (3.7 mm, V(
σ) 0.4 µL)
1 G4212-60038
ULD Max-Light Cartridge Cell (10 mm, V(
σ) 0.6 µL)
Preparations Turn the pump off.
Sample degradation and contamination of the instrument
Metal parts in the flow path can interact with the bio-molecules in the sample leading
to sample degradation and contamination.
➔ For bio-inert applications, always use dedicated bio-inert parts, which can be
identified by the bio-inert symbol or other markers described in this manual.
➔ Do not mix bio-inert and non-inert modules or parts in a bio-inert system.
The flow cell is shipped with a filling of isopropanol. This is to avoid breakage due to
subambient conditions. In case the flow cell is not used for some time (stored), then flush
the flow cell with iso-propanol.