Agilent 1200 Infinity Series DAD User Manual 167
Maintenance
9
Overview of Maintenance
Overview of Maintenance
The following pages describe maintenance (simple repairs) of the detector that
can be carried out without opening the main cover.
Tabl e 1 2 Overview of Maintenance
Procedure Typical Frequency Notes
Cleaning of module If required
Deuterium lamp
exchange
If noise and/or drift exceeds your application limits or
lamp does not ignite.
A wavelength calibration test and an
intensity test should be performed after
replacement.
Flow cell exchange If leaking or if intensity drops due to contaminated flow
cell.
A wavelength calibration test should be
performed after replacement.
Leak sensor drying If leak has occurred. Check for leaks.
Leak handling System
replacement
If broken or corroded. Check for leaks.