LAN Configuration
13
Dynamic Host Configuration Protocol (DHCP)
Dynamic Host Configuration Protocol (DHCP)
General Information (DHCP)
The Dynamic Host Configuration Protocol (DHCP) is an auto configuration
protocol used on IP networks. The DHCP functionality is available on all
Agilent HPLC modules with on-board LAN Interface or LAN Interface Card,
and “B”-firmware (B.06.40 or above).
When the initialization mode “DHCP” is selected, the card tries to download
the parameters from a DHCP Server. The parameters obtained become the
active parameters immediately. They are not stored to the non-volatile
memory of the card.
Besides requesting the network parameters, the card also submits its
hostname to the DHCP Server. The hostname equals the MAC address of the
card, e.g. 0030d3177321. It is the DHCP server's responsibility to forward the
hostname/address information to the Domain Name Server. The card does not
offer any services for hostname resolution (e.g. NetBIOS).
Figure 62 DHCP (Principle)
9=8E
HZgkZg
6Xi^kZ
EVgVbZiZg
NOTE
1 It may take some time until the DHCP server has updated the DNS server with the
hostname information.
2 It may be necessary to fully qualify the hostname with the DNS suffix, e.g.
0030d3177321.country.company.com.
3 The DHCP server may reject the hostname proposed by the card and assign a name
following local naming conventions.

Table of Contents

Other manuals for Agilent Technologies 1200 Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 Infinity Series and is the answer not in the manual?

Agilent Technologies 1200 Infinity Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 Infinity Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals