EasyManuals Logo

Agilent Technologies 1200 Infinity Series User Manual

Agilent Technologies 1200 Infinity Series
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #79 background imageLoading...
Page #79 background image
Agilent 1200 Infinity Series DAD User Manual 79
Optimizing the Detector
5
Optimization Overview
Optimization Overview
Tabl e 6 Optimization Overview
Parameter Impact
1 Selection of flow cell peak resolution versus sensitivity
Choose flow cell according to used column (“Choosing a Flow Cell” on page 80).
2 Connection of flow cell chromatographic resolution
3 Setting the peak width (response time) peak resolution versus sensitivity
versus disk space
Use peak width according “Choosing a Flow Cell” on page 80 as starting point.
Set the peak-width close to the width of a narrow peak of interest in your
chromatogram.
4 Setting wavelength and bandwidth
Sample wavelength:
Never miss a peak by the use of a browser wavelength like 250 nm with
100 nm bandwidth.
Select specific wavelength with reduced optical bandwidth if you need
selectivity, e.g. 254.0 nm / 4 nm and 360.0 nm / 100 nm as reference
wavelength.
Set the sample wavelength to a peak or valley to get best linearity in general;
select a valley to get best linearity for high concentrations.
sensitivity versus selectivity
sensitivity versus linearity
Reference wavelength:
Select the reference wavelength with broad bandwidth (30...100 nm)
wavelength range where your analytes have little or no absorbance
(e.g. sample at 254 nm, reference at 320 nm).
Select the reference wavelength as near as possible to the UV range.
baseline drift due to RI effects.
5 Setting the slit width (G4212A only)
Use 4 nm slit for normal applications.
Use narrow slit (e.g 1 nm) if your analytes have narrow absorbance bands and
for high concentrations.
Use a wide slit (e.g. 8 nm) to detect very low concentrations.
Optimizing spectral acquisition (DAD only)
Select spectra acquisition mode according to your needs (see “Spectrum
Settings” on page 61).
Set the spectral wavelength range (for colorless samples 190...400 nm is
sufficient).
Set step to 4 nm for normal use; set small step (and slit width) if high resolution
of spectra with fine structure is wanted.
spectral resolution, sensitivity and
linearity.

Table of Contents

Other manuals for Agilent Technologies 1200 Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 Infinity Series and is the answer not in the manual?

Agilent Technologies 1200 Infinity Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 Infinity Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals