28 Agilent 1200 Infinity Series DAD User Manual
2 Site Requirements and Specifications
Performance Specifications
Specification Conditions
ASTM: “Standard Practice for Variable Wavelength Photometric Detectors
Used in Liquid Chromatography”.
Reference conditions:
Wavelength: 230 nm/4 nm with Reference Wavelength 360 nm/100 nm,
Slitwidth 4 nm, TC 2 s, (or with RT = 2.2 * TC), ASTM
Max-Light Cartridge Cell ( 10 mm, V(σ) 1.0 μL) (G4212-60008) with flow of
0.5 mL/min LC grade water or Max-Light Cartridge Test Cell (G4212-60011)
Linearity:
Linearity is measured with caffeine at 265 nm/4 nm with slit width 4 nm and
TC 1 s (or with RT 2 s) with Max-Light Cartridge Cell ( 10 mm, V(σ) 1.0 μL)
(G4212-60008) > 2.0 AU (5 %) [ typical 2.5 AU (5 %) ].
ASTM drift tests require a temperature change below 2 °C/hour (3.6 F/hour)
over one hour period. Our published drift specification is based on these
conditions. Larger ambient temperature changes will result in larger drift.
Better drift performance depends on better control of the temperature
fluctuations. To realize the highest performance, minimize the frequency and
the amplitude of the temperature changes to below 1 °C/hour (1.8 F/hour).
Turbulences around one minute or less can be ignored.
Performance tests should be done with a completely warmed up optical unit (>
two hours). ASTM measurements require that the detector should be turned
on at least 24 h before start of testing.
Time Constant versus Response Time
According to ASTM E1657-98 „Standard Practice of Testing
Variable-Wavelength Photometric Detectors Used in Liquid Chromatography”
the time constant is converted to response time by multiplying by the factor
2.2.
NOTE
The specifications are based on the standard RFID tag lamp (5190-0917) and may be not
achieved when other lamp types or aged lamps are used.

Table of Contents

Other manuals for Agilent Technologies 1200 Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 Infinity Series and is the answer not in the manual?

Agilent Technologies 1200 Infinity Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 Infinity Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals