148 Agilent 1200 Infinity Series DAD User Manual
8 Test Functions and Calibration
ASTM Drift and Noise Test
ASTM Drift and Noise Test
The ASTM noise test determines the detector noise over a period of
20 minutes. The test is done with installed Max-Light Cartridge Cell or
Max-Light Cartridge Test Cell.
This test does also check for the drift. It is also part of the “Self Test” (without
checking for the drift).
If the test is performed with the Max-Light Cartridge Test Cell, the test results
are not influenced by solvent or pump effects.
When In case of noise and drift problem.
Parts required # Description
1 Max-Light Cartridge Cell (filled with water)
OR 1 Max-Light Cartridge Test Cell
Preparations Detector and UV-lamp must be on for at least 2 hours.
ASTM measurements based on specifications may require longer stabilization times.
When using a Max-Light Cartridge Cell a flow rate of 0.5 mL/min with water is required.

Table of Contents

Other manuals for Agilent Technologies 1200 Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 Infinity Series and is the answer not in the manual?

Agilent Technologies 1200 Infinity Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 Infinity Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals