48 Agilent 1200 Infinity Series DAD User Manual
3 Installing the Module
Initial Recalibration
Initial Recalibration
The detector has been calibrated with a factory flow cell initially. After
installation of the detector with the delivered or a new Max-Light Cartridge
flow cell and an initial warm-up of at least 2 hours, a recalibration should be
performed (“Wavelength Calibration” on page 156). This recalibration will
correct slight changes due to
significant environmental condition changes (temperature, humidity)
during transport and storage,
significant environmental condition changes (temperature, humidity) in
final location and
variances between factory test cell and the installed flow cell.