48 Agilent 1200 Infinity Series DAD User Manual
3 Installing the Module
Initial Recalibration
Initial Recalibration
The detector has been calibrated with a factory flow cell initially. After
installation of the detector with the delivered or a new Max-Light Cartridge
flow cell and an initial warm-up of at least 2 hours, a recalibration should be
performed (“Wavelength Calibration” on page 156). This recalibration will
correct slight changes due to
• significant environmental condition changes (temperature, humidity)
during transport and storage,
• significant environmental condition changes (temperature, humidity) in
final location and
• variances between factory test cell and the installed flow cell.