142 Agilent 1200 Infinity Series DAD User Manual
8 Test Functions and Calibration
Intensity Test
Test Failed
Intensity Test Evaluation
Probable cause Suggested actions
1 Absorbing solvent or air bubble in flow cell.
Ensure the flow cell is filled with water, and
free from air bubbles.
Repeat test with Max-Light Cartridge Test
Cell and compare results.
2 Incorrect calibration
Recalibrate and repeat the test.
3 Dirty or contaminated flow cell.
Run the cell test. If the test fails, flush the flow
cell. See also “Cleaning of Max-Light Cartridge
Cell” on page 178.
4 Dirty or contaminated optical components.
Please contact your Agilent service
representative.
5 Old UV-lamp.
Exchange the UV-lamp.
6 Defect optical unit.
If the test fails with Max-Light Cartridge Test
Cell and new UV-lamp, please contact your
Agilent service representative.
NOTE
If only one range fails and the application does not require this range, the lamp may not be
changed.

Table of Contents

Other manuals for Agilent Technologies 1200 Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 Infinity Series and is the answer not in the manual?

Agilent Technologies 1200 Infinity Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 Infinity Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals