98 Agilent 1200 Infinity Series DAD User Manual
5 Optimizing the Detector
Optimizing Selectivity
Ratio Qualifiers for Selective Detection of Compound Classes
Ratio qualifiers can be used where, in a complex sample, only one particular
class needs to be analyzed a parent drug and its metabolites in a biological
sample, for example. Another example is the selective analysis of derivatives
after pre- or post-column derivatization. Specifying a signal ratio that is
typical for the sample class is one way of selectively plotting only those peaks
that are of interest. The signal output remains at zero so long as the ratio is out
of the user-specified ratio range. When the ratio falls within the range, the
signal output corresponds to the normal absorbance, giving single, clear peaks
on a flat baseline. An example is shown in Figure 27 on page 98 and Figure 28
on page 99.
Figure 27 Wavelength Selection for Ratio Qualifiers
d"IZge]Zcna
7^e]Zcna
HZaZXi'X]VgVXiZg^hi^X
lVkZaZc\i]h
H^\cVa2LA&$LA''%
LVkZaZc\i]cb
LA&
LA'
HXVaZY

Table of Contents

Other manuals for Agilent Technologies 1200 Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 Infinity Series and is the answer not in the manual?

Agilent Technologies 1200 Infinity Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 Infinity Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals