Agilent 1200 Infinity Series DAD User Manual 27
Site Requirements and Specifications
2
Performance Specifications
Flow cells Max-Light Cartridge Cell ( 10 mm, V(σ) 1.0 µL) (G4212-60008),
Max-Light Cartridge Cell ( 60 mm, V(σ) 4.0 µL) (G4212-60007),
HDR Max-Light Cartridge Cell ( 3.7 mm, V(σ) 0.4 µL)
(G4212-60032)
ULD Max-Light Cartridge Cell ( 10 mm, V(σ) 0.6 µL)
(G4212-60038)
Max-Light Cartridge Cell Bio-inert ( 10 mm, V(σ) 1.0 µL)
(G5615-60018)
Max-Light Cartridge Cell Bio-inert ( 60 mm, V(σ) 4.0 µL)
(G5615-60017)
Max-Light Cartridge Test Cell (G4212-60011)
70 bar (1015 psi) Maximum
Operating Pressure (MOP)
1
150 bar (2175 psi) Maximum
Incidential Pressure (MIP)
2
pH range 1.0 —12.5 (solvent
dependent) available as standard
and bio-inert versions.
Cartridge type, equipped with
RFID tags that holds cell typical
information.
Local Control Agilent Instant Pilot (G4208A) B.02.11 or above
Tes t an d
troubleshooting
software
Agilent LabAdvisor B.01.03 SP4 or above
Analog outputs Recorder/integrator: 100 mV or 1 V, output range
0.001–2AU, one output
Communications Controller-area network (CAN), RS-232C, APG Remote: ready,
start, stop and shut-down signals, LAN
Safety and
maintenance
Extensive support for troubleshooting and maintenance is
provided by the Instant Pilot, Agilent Lab Advisor, and the
Chromatography Data System. Safety-related features are leak
detection, safe leak handling, leak output signal for shutdown
of pumping system, and low voltages in major maintenance
areas.
GLP features Early maintenance feedback (EMF) for continuous tracking of
instrument usage in terms of lamp burn time with user-setable
limits and feedback messages. Electronic records of
maintenance and errors. Verification of wavelength accuracy
with the emission lines of the deuterium lamp.
Housing All materials recyclable.
1
Maximum Operating Pressure (MOP): The maximum pressure at which the system can operate continuously under normal
conditions.
2
Maximum Incidential Pressure (MIP): The maximum pressure which the system can experience during a short time.
Tabl e 4 Performance Specifications G4212B
Type Specification Comments

Table of Contents

Other manuals for Agilent Technologies 1200 Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 Infinity Series and is the answer not in the manual?

Agilent Technologies 1200 Infinity Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 Infinity Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals