EasyManua.ls Logo

Agilent Technologies 1200 Infinity Series - Page 26

Agilent Technologies 1200 Infinity Series
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26 Agilent 1200 Infinity Series DAD User Manual
2 Site Requirements and Specifications
Performance Specifications
Performance Specifications G4212B
Tabl e 4 Performance Specifications G4212B
Type Specification Comments
Detection type 1024-element photodiode array
Light source Deuterium lamp Equipped with RFID tag that holds
lamp typical information.
Wavelength range 190 640 nm
Short term noise
(ASTM) Single and
Multi-Wavelength
< ± 3 × 10
-6
AU at 230 nm/4 nm, with 10 mm Max-Light
cartridge cell
Typically < ± 0.6 × 10
-6
AU/cm at 230 nm/4 nm, with 60 mm
Max-Light cartridge cell
see "Specification Conditions"
below
Drift
< 0.5 × 10
-3
AU/hr at 230 nm
see "Specification Conditions"
below
Linear absorbance
range
> 2.0 AU (5 %) at 265 nm see "Specification Conditions"
below
Wavelength
accuracy
± 1 nm After recalibration with
deuterium lines
Wavelength
bunching
2–400nm Programmable in steps of 1 nm
Slit width 4 nm Fixed slit
Diode width ~ 0.5 nm
Signal data rate 80 Hz
Spectra Data rate 80 Hz