EasyManua.ls Logo

Agilent Technologies 1200 Infinity Series - Page 157

Agilent Technologies 1200 Infinity Series
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1200 Infinity Series DAD User Manual 157
Test Functions and Calibration
8
Wavelength Calibration
1 Run the Wavelength Calibration with the recommended user interface (for
further information see Online-Help of user interface).
Figure 46 Wavelength Calibration – Results
NOTE
If the detector is operated in a lab environment that differs at average from the final test
environment (25 °C) then the detector should be recalibrated for this temperature.
NOTE
If the detector was repaired (opened covers), the wavelength calibration can be done
10 minutes after lamp on. A final wavelength calibration should be repeated after complete
warm-up of the detector.

Table of Contents

Other manuals for Agilent Technologies 1200 Infinity Series

Related product manuals