Agilent 1200 Infinity Series DAD User Manual 157
Test Functions and Calibration
8
Wavelength Calibration
1 Run the Wavelength Calibration with the recommended user interface (for
further information see Online-Help of user interface).
Figure 46 Wavelength Calibration – Results
NOTE
If the detector is operated in a lab environment that differs at average from the final test
environment (25 °C) then the detector should be recalibrated for this temperature.
NOTE
If the detector was repaired (opened covers), the wavelength calibration can be done
10 minutes after lamp on. A final wavelength calibration should be repeated after complete
warm-up of the detector.