Agilent 1200 Infinity Series DAD User Manual 229
LAN Configuration
13
Initialization mode selection
Using Stored
When initialization mode Using Stored is selected, the parameters are taken
from the non-volatile memory of the module. The TCP/IP connection will be
established using these parameters. The parameters were configured
previously by one of the described methods.
Figure 60 Using Stored (Principle)
Using Default
When Using Default is selected, the factory default parameters are taken
instead. These parameters enable a TCP/IP connection to the LAN interface
without further configuration, see Table 23 on page 229.
Figure 61 Using Default (Principle)
Cdc"KdaVi^aZ
G6B
HidgZY
EVgVbZiZg
6Xi^kZ
EVgVbZiZg
6Xi^kZ
EVgVbZiZg
9Z[Vjai
EVgVbZiZg
NOTE
Using the default address in your local area network may result in network problems. Take
care and change it to a valid address immediately.
Tabl e 2 3 Using Default Parameters
IP address: 192.168.254.11
Subnet Mask: 255.255.255.0
Default Gateway not specified