Agilent 1200 Infinity Series DAD User Manual 59
Using the Module
4
Setting up the Detector with Agilent ChemStation
Peakwidth (time programmed)
These selections can be made during time programmed
operation.
When used in a timetable, Peakwidth changes the filters
used for peak-controlled spectra acquisition, but not the
data rate of a chromatographic signal.
NOTE
This setting makes sense only with peak-controlled
spectra; it allows you to change the peakwidth setting
to account for broadening peaks at the end of the run.
Stoptime / Posttime
The stoptime is the time where either the complete system
stops (As Pump/Injector) or the module (if different from
system stop time). The data collection is stopped at this
time.
A posttime period can be used to allow module’s items to
equilibrate (e.g. after gradient change or temperature
change).