EasyManua.ls Logo

Agilent Technologies 1200 Infinity Series - Page 59

Agilent Technologies 1200 Infinity Series
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1200 Infinity Series DAD User Manual 59
Using the Module
4
Setting up the Detector with Agilent ChemStation
Peakwidth (time programmed)
These selections can be made during time programmed
operation.
When used in a timetable, Peakwidth changes the filters
used for peak-controlled spectra acquisition, but not the
data rate of a chromatographic signal.
NOTE
This setting makes sense only with peak-controlled
spectra; it allows you to change the peakwidth setting
to account for broadening peaks at the end of the run.
Stoptime / Posttime
The stoptime is the time where either the complete system
stops (As Pump/Injector) or the module (if different from
system stop time). The data collection is stopped at this
time.
A posttime period can be used to allow module’s items to
equilibrate (e.g. after gradient change or temperature
change).

Table of Contents

Other manuals for Agilent Technologies 1200 Infinity Series

Related product manuals