Agilent 1200 Infinity Series DAD User Manual 81
Optimizing the Detector
5
Optimization Overview
Ultra-Low Dispersion
The Max-Light Cartridge ULD cell can be used with the G4212A DAD and
G4212B DAD. The cell is a requirement for the Ultra-Low Dispersion Kit
solution which currently exists as 1290 Infinity Ultra-Low Dispersion Kit
(5067-5189). The cell should be part of the ultra-low dispersion solution.
High Dynamic Range
The Max-Light Cartridge HDR cell can be used with the G4212A DAD and
G4212B DAD. The cell is required as a part of the High Dynamic Range (HDR)
solution which will be introduced March/April 2013.
Recommendations
For G4212-60007 and G4212-60008
The use of Peek-FS capillaries is not recommended. In combination with the
SST zero dead volume fitting (e.g. at the inlet) the capillary could break and
the glass particles could block/damage the flow cell.
NOTE
To protect the flow cell against overpressure (e.g. in systems with LC/MS) install Inline
Pressure Relief Valve Kit (G4212-68001), see “Inline Pressure Relief Valve Kit
(G4212-68001)” on page 82.

Table of Contents

Other manuals for Agilent Technologies 1200 Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 Infinity Series and is the answer not in the manual?

Agilent Technologies 1200 Infinity Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 Infinity Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals