Agilent 1200 Infinity Series DAD User Manual 81
Optimizing the Detector
5
Optimization Overview
Ultra-Low Dispersion
The Max-Light Cartridge ULD cell can be used with the G4212A DAD and
G4212B DAD. The cell is a requirement for the Ultra-Low Dispersion Kit
solution which currently exists as 1290 Infinity Ultra-Low Dispersion Kit
(5067-5189). The cell should be part of the ultra-low dispersion solution.
High Dynamic Range
The Max-Light Cartridge HDR cell can be used with the G4212A DAD and
G4212B DAD. The cell is required as a part of the High Dynamic Range (HDR)
solution which will be introduced March/April 2013.
Recommendations
For G4212-60007 and G4212-60008
The use of Peek-FS capillaries is not recommended. In combination with the
SST zero dead volume fitting (e.g. at the inlet) the capillary could break and
the glass particles could block/damage the flow cell.
To protect the flow cell against overpressure (e.g. in systems with LC/MS) install Inline
Pressure Relief Valve Kit (G4212-68001), see “Inline Pressure Relief Valve Kit
(G4212-68001)” on page 82.