Agilent 1200 Infinity Series DAD User Manual 91
Optimizing the Detector
5
Optimizing for Sensitivity, Selectivity, Linearity and Dispersion
Because the detector averages absorbance values that are calculated for each
wavelength, using a wide bandwidth does not negatively impact linearity.
The use of a reference wavelength is highly recommended to further reduce
baseline drift and wander induced by room temperature fluctuations or
refractive index changes during a gradient.
An example of the reduction of baseline drifts is shown in Figure 22 on
page 91 for PTH-amino acids. Without a reference wavelength, the
chromatogram drifts downwards due to refractive index changes induced by
the gradient. This is almost completely eliminated by using a reference
wavelength. With this technique, PTH-amino acids can be quantified in the low
picomole range even in a gradient analysis.
Figure 22 Gradient Analysis of PTH-Amino Acids (1 pmol each), with and without
Reference
I^bZb^c
<gVY#/%#%'b@=ED$68C[gdb&'68Cid)*68C^c&'b^c
&ebdaZVX]
GZ[ZgZcXZ(-%cb
LVkZaZc\i]
'+,cb
LVkZaZc\i]
'+,cb
CdgZ[ZgZcXZ
EI="6HC
EI="6G<
EI="6A6
EI="E=:
EI="EGD
')

Table of Contents

Other manuals for Agilent Technologies 1200 Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 Infinity Series and is the answer not in the manual?

Agilent Technologies 1200 Infinity Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 Infinity Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals