CALCulate[1|2|3|4]:LIMit:UPPer:SEGMent:CLEar command
Deletes a segment from the upper limit line.
Command Syntax: CALCulate[1|2|3|4]:LIMit:UPPer:SEGMent:CLEar <number>|
<bound>
<number> ::= a real number (NRf data)
limits: -9.9e37:9.9e37
(depend upon current range of the X-axis)
<bound> ::= MAX|MIN
Example Statements: OUTPUT 711;"calculate4:lim:upp:segment:cle -3.21429e+37"
OUTPUT 711;"Calculate4:Lim:Upp:Segment:Cle -2.22796e+37"
Attribute Summary: Option: not applicable
Synchronization Required: no
Preset State: not applicable
SCPI Compliance: instrument-specific
Description:
This command deletes any segment which contains the X-axis value you specify. Adjacent segments are
not affected, although the limit line may be discontinuous.
The value entered for a limit line segment is unitless.
To delete all segments of an upper limit line, send CALC:LIM:UPP:CLE.
CALCulate
6-50