Chapter 5 257
Post-Repair Procedures
Baseband I/Q Amplitude Accuracy Test
I only
600 Ω Balanced
a
1 I input −22 ± 1.0
I only 1 MΩ Unbalanced 1 I input −10 ± 1.0
I only
1 MΩ Balanced
a
1 I input −10 ± 1.0
I only
1 MΩ Balanced
a
1I input (I bar) −10 ± 1.0
Q only 50 Ω Unbalanced 1 Q input −16 ± 1.0
Q only 50 Ω Unbalanced 0.25 Q input −16 ± 1.0
Q only 50 Ω Unbalanced 0.125 Q input −16 ± 1.0
Q only
600 Ω Balanced
a
1Q input −22 ± 1.0
Q only 1 MΩ Unbalanced 1 Q input −10 ± 1.0
Q only
1 MΩ Balanced
a
1Q input −10 ± 1.0
Q only
1 MΩ Balanced
a
1Q input (Q bar) −10 ± 1.0
a. Whenever performing a balanced measurement ensure the complimentary input
is terminated (for a balanced input, when the signal source is connected to the I
input, the I
input must be terminated).
Table 5-8 Amplitude Accuracy at 100 kHz and 5 MHz
Input
Port
I/Q Input Z I/Q
Range
(V)
I/Q Input
Connector
UUT Measured
Amplitude
(dBm)
Nominal
Amplitude
(dBm)
Test
Limits
(dB)
100 kHz 5 MHz

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals