Chapter 5 257
Post-Repair Procedures
Baseband I/Q Amplitude Accuracy Test
I only
600 Balanced
a
1 I input 22 ± 1.0
I only 1 M Unbalanced 1 I input 10 ± 1.0
I only
1 M Balanced
a
1 I input 10 ± 1.0
I only
1 M Balanced
a
1I input (I bar) 10 ± 1.0
Q only 50 Unbalanced 1 Q input 16 ± 1.0
Q only 50 Unbalanced 0.25 Q input 16 ± 1.0
Q only 50 Unbalanced 0.125 Q input 16 ± 1.0
Q only
600 Balanced
a
1Q input 22 ± 1.0
Q only 1 M Unbalanced 1 Q input 10 ± 1.0
Q only
1 M Balanced
a
1Q input 10 ± 1.0
Q only
1 M Balanced
a
1Q input (Q bar) 10 ± 1.0
a. Whenever performing a balanced measurement ensure the complimentary input
is terminated (for a balanced input, when the signal source is connected to the I
input, the I
input must be terminated).
Table 5-8 Amplitude Accuracy at 100 kHz and 5 MHz
Input
Port
I/Q Input Z I/Q
Range
(V)
I/Q Input
Connector
UUT Measured
Amplitude
(dBm)
Nominal
Amplitude
(dBm)
Test
Limits
(dB)
100 kHz 5 MHz