EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>FieldFox N9913A

Agilent Technologies FieldFox N9913A User Manual

Agilent Technologies FieldFox N9913A
219 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #138 background imageLoading...
Page #138 background image
138 FieldFox User’s Guide
Display Units
Select whether to display power in dBm or Watts. When making relative power
measurements, these change to dB and %.
How to set Power Meter display units
Press Meas Setup 4 or Scale / Amptd.
Then Unit .
Then choose from the following:
o dBm (dB) Sets Power Meter display to dBm or dB.
o Watt (%) Sets Power Meter display to Watts or %.
Resolution
Resolution sets the number of digits to display after the decimal point on the
digital power meter display.
How to set Resolution
Press Scale / Amptd.
Then More.
Then choose from 0, 1, 2, or 3.
Averaging
Averaging is used to improve measurement accuracy in low power or noisy
power situations. Increased averaging improves measurement accuracy, but also
reduces measurement speed.
How to set Power Meter Averaging
Press BW 2 or Meas Setup 4.
Then choose from the following:
o Averaging Auto Man Off
o Auto Allows the USB Power Sensor to use its Auto Averaging settings.
See the USB Power Sensor documentation for details.
o Man Manually enter an averaging count by setting Num Averages.
o OFF Performs NO averaging.
o Num Averages Sets the number of Power Meter readings to be averaged
before a valid reading is displayed. When a number is entered, Averaging is
automatically set to Man.
Single or Continuous Measure
This setting determines whether the FieldFox measures continuously or only
once each time the Single button is pressed. Use Single to conserve battery
power or to allow you to save or analyze a specific measurement.
This setting can be changed at any time without affecting calibration accuracy.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies FieldFox N9913A and is the answer not in the manual?

Agilent Technologies FieldFox N9913A Specifications

General IconGeneral
BrandAgilent Technologies
ModelFieldFox N9913A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals