MM-400A
8. Operation Screens
8-105
(3) WELD COUNTER PRESET
Set a preset count value (0 to 999,999) of the weld counter. The weld counter
counts up by 1 for each measurement whether the value is within upper and
lower limits or not. When the counter reaches the preset count value, the
COUNT UP signal is output. When “0” is set for the preset count value, the
COUNT UP signal is not output.
(4) GOOD COUNTER PRESET
Set a preset count value (0 to 999,999) of the good counter. The good counter
counts up by 1 for each measurement only when the value is within upper and
lower limits. When the counter reaches the preset count value, the COUNT UP
signal is output. When “0” is set for the preset count value, the COUNT UP
signal is not output.
(5) DATE AND TIME
Set the date in the format: year (2016 to 2077), month (1 to 12) and day (1 to
31). Set the time in the format: hour (0 to 23) and minutes (0 to 59).
(6) BRIGHTNESS
Set the brightness of a screen in the range of 01 to 10 (01: dark ... 10: bright).
When AUTO is set, the energy saving mode is started if no operation is
performed within continuing three minutes, and the screen display disappears
automatically. Touching the screen again redisplays the screen.
(7) Function keys
MENU: Touching this displays the MENU screen.
NEXT: Touching this displays the SYSTEM SETUP (2) screen.
CT RESET: Touching this resets the counter value.
PASSWRD: Touching this displays the PASSWORD screen.
INIT: Touching this initializes all setting conditions and envelope conditions. It
takes about 60 seconds for initialization.
IMPORTANT
Do not turn off the power supply during initialization. If not, it results in
malfunction.