If the SWEEP group is selected, the screen will appear as shown in Figure 9-24.  The screen 
has the following parameters that are unique to the group: 
 
Figure 9-24: IEC 1000-4-13 SWEEP 
1.  LEVEL   Sets the percentage level of the interharmonics relative to the fundamental. 
     The level is fixed for the entire frequency range, which is defined by 
     FRANGE. To change level, the test state must be off and user class must be 
     selected. 
2.  FRANGE      The range is selected by rotating the shuttle. The range selection is required 
     to change the level for the selected range.  Also, an individual range sweep 
     is possible by selecting the desired range and selecting RUN RANGE from 
     the RUN field. 
3.  STEP      The step size defines the interharmonics sweep points.  Step size is fixed for 
     the entire sweep range.  The STEP size can be changed when the test state 
     is off. 
4.  IHFREQ   This field will show the current interharmonic frequency when the test is 
     running.  An interharmonic frequency can be selected using the shuttle.  To 
     run a frequency without sweeping through the entire range, a RUN POINT 
     can be selected. 
5.  RESONANT   This field is used to report the resonant points, if any, after running the sweep 
     test.  No editing is allowed in this field.  When the test is completed, a display 
     of harmonics current versus frequency plot is available.  To view the graph, 
     select the field and press the ENTER key. 
HARMONICS GROUP 
If the Harmonics group is selected, the screen will appear as shown in Figure 9-25. The screen 
has the following parameters that are unique to the group: 
 
Figure 9-25: IEC 1000-4-13 Harmonics 
1.  LEVEL  Sets the percentage level of the harmonic relative to the fundamental.  Each 
    level is associated with a harmonic number. To change levels, the test state 
    must be off and user class must be selected. 
2.  HARM NO  Selecting a harmonic number using the shuttle knob will allow examination of 
    the LEVEL and PHASE of each harmonic and changing the level and phase 
    angle when USER class is selected. 
3.  PHASE  Sets the phase angle of the harmonics relative to the fundamental.  To change 
    phase, the test state must be off and user class must be selected. 
4.   PAUSE  Sets the pause time in seconds between harmonics tests.  There is only one 
    value for the entire test. The PAUSE value can be changed when the test is not 
    running.