User Manual    AMETEK Programmable Power 
RS Series    147 
 
6.3.3  Measurement Calibration Summary 
The following Table is a summary of the preceding calibration steps. The value indicated by the 
External DVM is called V
AC
 or V
DC
.  The current measured by the current shunt is called I
AC
 or 
I
DC
. 
 
TITLE  PROGRAM/LOAD PARAMETERS  PARAMETER  ADJUST TO 
AC MODE       
AC Volt Full-scale  300 VAC Range, 240 VAC, 60 Hz, no load  VOLT FS  V
AC
 
AC Current Full-scale  150 VAC Range, 120 VAC, 60 Hz, full load 
to 90% of max current range. 
CURR FS  I
AC
 
DC MODE       
DC Volt + Full-scale  400 VDC Range, + 320 VDC, no load  VOLT FS  V
DC
 
DC Current Full-scale  200 VDC Range, 160 VDC, full load to 90% 
of max. current range. 
CURR FS  I
DC
 
Table 6-2: Measurement Calibration Table - TBD 
 
Repeat Paragraph 6.3 for each phase.  Move the external test equipment to the phase that is 
being calibrated.  Refer to Figure 6-2.. 
While viewing the calibration screen, press the PHASE key to select the respective phase.