9.5 APR [Application Result]
Type 3: LED measurement
Value of λ fwhm
λ fwhm indicates the center wavelength obtained through the RMS analysis (k =
2.35).
The unit is always nm. A value is output down to the third decimal place.
When analysis is impossible, –1 is output.
Value of λ ndb
λ ndb indicates the center wavelength obtained through the threshold analysis
(ndB).
The unit is always nm. A value is output down to the third decimal place.
When analysis is impossible, –1 is output.
Value of FWHM
FWHM indicates the half-magnitude full width obtained through the RMS analy-
sis (k = 2.35)
The unit is always nm. A value is output down to the third decimal place.
When analysis is impossible, –1 is output.
Value of BWndb
BWndb indicates the wave width obtained through the threshold analysis (ndB).
The unit is always nm. A value is output down to the third decimal place.
When analysis is impossible, –1 is output.
Value of λ p
λ p indicates a peak wavelength.
The unit is always nm. A value is output down to the third decimal place.
When analysis is impossible, –1 is output.
Value of Lp
Lp indicates a peak level.
The unit is dBm. A value is output down to the second decimal place.
When analysis is impossible, –999.99 is output.
Value of PKdens
PKdens indicates a peak power per mm.
The unit is always dBm. A value is output down to the second decimal place.
When analysis is impossible, –999.99 is output.
Value of POW
POW indicates a power integration value.
The unit is always dBm. A value is output down to the second decimal place.
When analysis is impossible, –999.99 is output.