EasyManua.ls Logo

Anritsu MS9710C - Status Messages

Anritsu MS9710C
269 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 3 Standards
3-6
3.3.2 Status Messages
Shown below is the structure of the service request summary message set in the status byte register of the MS9710C.
Summary Bit Configuration of Status Byte Register
7
6
5
4
3
2
1
0
7
6
5
4
3
2
1
0
Power-on
User request
Command error
Execution error
Device-dependent error
Query error
Bus control request
Operation complete
&
&
&
&
&
&
&
&
Logical OR
Bit 7
DIO8
Reserved
Bit
Line
Summary
message bit
Bit 6
DIO7
Bit 5
DIO6
ESB
Bit 4
DIO5
MAV
Bit 3
DIO4
ERR
Data
Data
Data
Data
Data
Data
Bit 2
DIO3
END
Bit 1
DIO2
Reserved
Bit 0
DIO1
Reserved
R
Q
S
M
S
S
Summary bit from next
page ESB (ERROR)
Summary bit from
next page ESB (END)
Output queue
Event Summary Bit
(ESB)
MAV summary bit indicating
the output queue is not empty
Message Available (MAV)
Standard event status
register
Standard event status
enable register
bit 0 to 5, 7
Service Request
Generation
Service Request
Enable Request
bit 0 to 5, 7
&
of
corresponding
bits

Table of Contents

Related product manuals