EasyManuals Logo

Anritsu MS9710C User Manual

Anritsu MS9710C
269 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #165 background imageLoading...
Page #165 background image
Section 9 Details on Device Messages
9-26
9.9 AUT [Auto Measure]
Function
Carries out auto measurement. The wavelength and resolution are automatically
set for the incident light spectrum.
When the measurement is complete, bit 0 (measurement end bit) of the extended
event status register (ESR2) is set to 1.
Header Program Query Response
AUT AUT AUT ? n
Value of n
n indicates a measurement state. 0 or 1 is output.
0: Measurement has been completed.
1: Measurement is being carried out.
9.10 AVS [Sweep Average]
Function
Sets the number of sweep operations for sweep averaging.
Header Program Query Response
AVS AVS n AVS ? n
Value of n
The unit is number of times. Input an integer comprising 1 to 3 digits or OFF.
Data range: 2 n 1000 , or OFF
Initial setting value
n is a backed up value.
Default
n = OFF

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu MS9710C and is the answer not in the manual?

Anritsu MS9710C Specifications

General IconGeneral
Frequency Range600 to 1750 nm
Wavelength Range600 to 1750 nm
Display5.7-inch color LCD
InterfacesGPIB, USB, Ethernet
Input Power+20 dBm
Battery LifeN/A (AC power only)