Section 9 Details on Device Messages
9-66
9.93 WDP [Vacuum/Air Set]
■ Function
Sets the wavelength display mode to “in Vacuum” or “in Air.”
Header Program Query Response
WDP WDP s WDP ? s
■ Value of s
s = VACUUM: in Vacuum
= AIR: in Air
■ Initial setting value
s is a backed up value.
■ Default
s = AIR
9.94 WOFS [Wavelength Offset]
■ Function
Sets a wavelength offset value.
Header Program Query Response
WOFS WOFS n WOFS ? n
■ Value of n
n is a waveform offset value.
The unit is always nm. Input a value down to the second decimal place.
Data range: –1.00 nm ≤ n ≤ 1.00 nm