Section 9 Details on Device Messages
9-60
9.80 TDL [EXT-trigger Delay Time]
■ Function
Sets a trigger delay used for EXT trigger measurement.
Header Program Query Response
TDL TDL n TDL ? n
■ Value of n
The unit is always m s. Input a value with an integer comprising 1 to 7 digits.
Data range: 0 ≤ n ≤ 5000000
■ Initial setting value
n is a backed up value.
■ Default
n = 0
9.81 TDSP [Time & Date Display On/Off]
■ Function
Turns on/off the timer display.
Header Program Query Response
TDSP TDSP s TDSP ? s
■ • Value of s
S indicates whether the timer display is on or off.
s = ON: Turn on the timer display.
= OFF:Turn off the timer display.
9.82 TER [Title Erase]
■ Function
Clears all title characters.
Header Program Query Response
TER TER None None