9-51
9.58 MPT [Sampling (Measureing) Points]
Function
Sets a number of sampling points.
Header Program Query Response
MPT MPT n MPT ? n
Value of n
The unit is number of points. Input one of the following values:
51, 101, 251, 501, 1001, 2001, 5001
Initial setting value
n is a backed up value.
Default
n = 501 (points)
9.59 MSL [Memory Select]
Function
Selects memory A or B for saving measurement data.
Header Program Query Response
MSL MSL s MSL ? s
Value of s
s = A: Select memory A.
= B: Select memory B.
Initial setting value
s = A (memory A)
9.60 OPT [Light Output]
Function
Turns on/off the light output from the internal light source (optional white light
source or reference light source).
Header Program Query Response
OPT OPT s OPT ? s
Value of n
s = ON: Turn on the light output.
= OFF:Turn off the light output.