9-49
9.55 MKD [Level Marker D]
9.55 MKD [Level Marker D]
Function
Sets level marker D with a level.
Header Program Query Response
MKD MKD l MKD ? l
Value of l
λ is a level value including the unit of the trace marker on the set scale.
When the log scale is selected, λ is a value represented down to the third decimal
place. When the linear scale is selected, it is represented by a maximum of seven
digits.
Data range:
–190 DBM λ ≤ +50 DBM ....Log scale, normal/max. hold/overlap, trace A/B/AB
–160 DB λ ≤ +160 DB......Log scale, normal/max. hold/overlap, trace A-B/
B-A
Log scale, normalize
–200 DB λ ≤ +120 DB......Linear scale, normal/max. hold/overlap, trace A/
B/AB
0 PW λ ≤ 1.2 W..........Linear scale, normal/max. hold/overlap, trace A/
B/AB
0 PCT λ ≤ 240 PCT ......Linear scale, normal/max. hold/overlap, trace A-
B/B-A
Linear scale, normalize
Suffixes
Same as level marker C.