9-55
9.67 PWRR [Power Monitor Result]
Function
Reads the result of measurement made using a power monitor.
Header Program Query Response
PWRR None PWRR ? p
Response
Value of p
p is a measured power value.
The unit is dBm. A value is output down to the second decimal place.
Data is tail-zero-suppressed.
9.68 RCL [FD File Recall]
Function
Recalls the specified file on the FD.
When file recall is complete, bit 2 (transfer end bit) of the extended event status
register (ESR2) is set to 1.
Header Program Query Response
RCL RCL n None None
Value of n
Input a DOS-recognizable file name comprising eight or fewer characters in the
following format:
xxxxxxxx.dat
The file extension “.dat” may be omitted.
9.68 RCL [FD File Recall]