9.18 DATE [Date Set]
Function
Sets the year, month, and day.
Header Program Query Response
DATE DATE yy,mm,dd DATE ? yy,mm,dd
• Value of yy
Input the last two digits (00 to 99) of Gregorian year.
Value of mm
Input two digits (01 to 12) of month.
Value of dd
Input two digits (01 to 31) of day.
9.19 DBA [Memory Data A]
Function
Outputs binary measurement data equivalent to the number of sampling points
from memory A.
The count of data in memory A is equal to the number of sampling points.
Header Program Query Response
DBA None DBA ? d
(equivalent to the number of
sampling points; binary data)
Value of d
d indicates measurement data. The data format depends on the scale.
See Appendix B, “Binary Data Transfer Format.”
Note:
In the 3-dimensional or overlapping display mode, the last measurement
data is output.
In the normalize display mode, an absolute level is output.